Medjil User Guides for Survey Instrument Calibration



Please login or signup to manage and calibrate your staff.

or

EDMI Calibration User Guide

  1. 1. Introduction
  2. 1. Background
  3. 3. Controlling Legislation
  4. 4. EDM Baselines provided in Western Australia
    1. 4.1 Curtin University Baseline
    2. 4.2 Kalgoorlie Baseline
    3. 4.3 Busselton Baseline
  5. 5. Instrument Calibration Procedure
    1. 5.1 Booking sheets
    2. 5.2 Preparation of equipment
    3. 5.3 Set-up and shade
    4. 5.4 Atmospheric correction control
    5. 5.5 Height of instrument
    6. 5.6 Reflector mountings
    7. 5.7 Levelling of equipment
    8. 5.8 Conditions
    9. 5.9 Calibration measurements
  6. 6. Measurement Sequence
    1. 6.1 Curtin University Baseline
    2. 6.2 Kalgoorlie Baseline
    3. 6.3 Busselton Baseline
  7. 7. Medjil Procedures
  8. 8. Potential error sources
  9. 9. Reduction and Interpretation
  10. 10. Definitions

1. Introduction

This guide outlines the calibraion of Electronic Distance Meters (EDM) in Western Australia.

There are three EDM instrumnt calibration baselines in the state that are maintained as subsidiary standards of length and as such are suitable for the calibration and standardisation of EDM instruments. All baselines are designed to support the mass of EDM unit lengths available in the modern market.

  • Curtin University, Kent St, Bentley, 24hrs 7days a week
  • Kalgoorlie Explosives Reserve, Piccadilly St, 7am to 3pm, Mon-Fri. Booking required
  • Busselton Bypass 24hrs 7days a week

Landgate re-calibrates EDM baselines every two years or as required. Check the EDM Baseline Regulation 13 certificate for details.

EDM instrument operators who wish to perform calibrations need to follow the procedures in this document to attain an acceptable calibration of their equipment.

The calibration of EDM instrumentation is concerned with the determination of instrument errors. Standardisation refers to the comparison of the instrument to a standard of length traceable to the national standard. Instruments must be calibrated within a prescribed level of precision to be standardised.

It is beyond the scope of this procedure to identify all sources of error inherent in EDM equipment. Surveyors should be aware of the limitations of the equipment they use and ensure that it is well maintained and regularly checked by the manufacturer’s agent.

2. Background

In 1983, the National Standards Commission (NSC), now incorporated into the National Measurement Institute (NMI), formed a working party on the Calibration of Electromagnetic Distance Measuring (EDM) Equipment. Following both the recommendations of this working party and research by the NSC, it was established that monumented baselines could be certified as subsidiary standards of length under Regulation 13 of the National Measurement Regulations 1999 to provide legal traceability for EDM measurements.

Recommendations of specific interest from the NSC working party on the calibration of EDM Equipment of 1 February 1983, include:

  • No.2 To be certified as a subsidiary standard a baseline must be capable of being calibrated with an uncertainty of ±[1.5 + (20 x 10-3 x L)] mm at the 95% level of confidence where L is the interval length in metres.
  • No.8 It is recommended that, in general, the minimum standard for the uncertainty of calibration of an EDM, assuming calibration against a monumented base, should be ±[4 + (20 x 10-3 x L)] mm at the 95% level of confidence where L is the interval length in metres.

In accordance with Regulation 73 of the National Measurement Regulations 1999, Surveyor-General of Western Australia is appointed as a Verifying Authority with respect to length. This enables certification of subsidiary standards of length to a certain precision pursuant to Regulation 13 of the National Measurement Regulations 1999.

Three Western Australian EDM baselines are certified biennial by the Surveyor-General as a reference standard under Regulation 13 of the National Measurement Regulations 1999 (Cth). The calibration of the EDM Baseline is certified annually in accordance with Recommendation No.2.

The calibration procedures outlined in this guide and the analysis techniques contained in the calibration portal are capable of meeting the requirement of Recommendation No.8.

3. Controlling Legislation

Matters relating to the uncertainty of surveys are detailed in Regulation 5 of the Licensed Surveyors (General Surveying Practice) Regulation 1961.

Matters relating to EDMI calibration are detailed in Regulation 20 of the Licensed Surveyors (General Surveying Practice) Regulation 1961.

4. EDM Baselines provided in Western Australia

    4.1 Curtin University Baseline

    In 1995 the Surveyor General reconditioned a ten-pillar baseline at Curtin University which had been in use for approximately ten years. Renovation included the placement of two additional pillars making the baseline a twelve-pillar baseline. New stainless steel pillar plate tops were provided to all pillars, with a 5/8" Whitworth threaded bolt at the centre of each pillar top. This baseline is situated on Curtin University campus parallel to Kent Street, Bentley. The baseline is partially accessible by vehicles at all times via the track alongside the baseline. Permission is not necessary to use the baseline. Any conflict of usage should be decided on a first to occupy basis.

    4.2 Kalgoorlie Baseline

    This EDM calibration baseline is situated on the Department of Mines and Petroleum Resources Explosives Reserve in Piccadilly Street, Kalgoorlie. The Kalgoorlie baseline consists of eight pillars set out such that all pillars are colinear. The baseline is readily accessible to 2WD vehicles when dry and is situated _1 a non-volatile area of the reserve. Phone in advance to ensure access 9091 7590. Any conflict of usage should be decided on a first to occupy basis.

    4.3 Busselton Baseline

    This EDM calibration baseline is situated on the Busselton Bypass west of Redgum Way and east of the Bussell Highway. The baseline consists of six pillars set out such that all pillars are colinear. The baseline is readily accessible to all vehicles. The baseline is accessible by vehicles at all times via the track alongside the length of the baseline. Permission is not necessary to use the baseline. Any conflict of usage should be decided on a first to occupy basis.

5. Instrument Calibration Procedure

    5.1 Booking sheets

    It is recommended that calibration data be recorded digitally and on official booking sheets provided by Landgate. All details should be recorded and booking sheets signed and dated. Measurements shall be recorded in units of metres (distance), degrees Celsius (temperature) and millibars (pressure). Baseline users are required to provide their own copies of booking sheets for use. Booking sheet may be downloaded from the Landgate website.

    5.2 Preparation of equipment

    Check the levelling bubbles on all tribrachs, reflectors and the theodolite, and if necessary, adjust before observing distances. Complete EDM instrument 'check and adjust' as per manufacturer instructions.

    5.3 Set-up and shade

    The instrument must be shaded by an umbrella at all times during the calibration. At no time should it be put in its box or left in the sun. It must be switched on and allowed to run, in the shade, for at least 15 minutes before measurements commence. The instrument should remain switched on during the whole calibration process.

    5.4 Atmospheric correction control

    For instruments with a phase measurement type, it is recommended that the atmospheric correction for the EDM be set to zero (ppm).

    For instruments with a pulse measurement type, the meteorological observations must be entered into the instrument at the time of the calibration. Distance readings will then be first-velocity corrected by the on-board EDM software.

    Ensure meteorological equipment has stabilised before making observations. When meteorological observations are entered into the instrument for application by the on-board EDM software, it is essential that any corrections to the meteorological observations are applied before inputting into the instrument.

    5.5 Height of instrument

    The height of the mounted EDM above the base of the tribrach (pillar plate) must be measured accurately to 1mm. This should be done with the foot screws in mid-setting.

    5.6 Reflector mountings

    The same reflector, reflector mounting, and tribrach should be used for all measurements. The height of the reflector must be measured and recorded in the same manner as for the EDM. The reflector must have a unique identification (serial number), which must be entered on the booking sheet.

    5.7 Levelling of equipment

    All equipment should be levelled with care on each stations. Start with all foot screws in the mid position. The stainless steel pillar tops have been setlevel on pillars.

    5.8 Conditions

    All calibration measurements must be taken either fully in daytime or fully at night. A mixture of conditions is not acceptable. EDM that are typically used in daytime should be calibrated in daytime.

    5.9 Calibration measurements

    On each line, four separate distance measurements should be taken as a minimum, with re-pointing after each measurement. Pointing can be optically or electronically performed as prescribed by the manufacturers.

6. Measurement Sequence

The Western Australian baselines were designed and constructed at a time when the majority of EDM in use had either 10m or 20m unit lengths. The modern trend in EDM, however, is smaller unit lengths, with 1.5m, 2m, 3m and 5m instruments being more common. There are also EDM in use with unit lengths of 30m and 33m. Medjil has compiled a list EDMs and their associated unit lengths that are presented as recommmended specification for instrument make and models.

It is recommended for all instruments, regardless of unit lengths, that an increased number of measurements be taken. The instrument correction determined from these measurements will satisfy the requirements of the NMI. The additional measurements will strengthen the calibration accuracy and reliability, and the additional redundancy will support any potential calibration issues associated with pillar movement or erroneous measurements.

    6.1 Curtin University Baseline

    • Pillar 1 at this site is no-longer suitable for use due to the line-of-sight obstruction by the tree between pillar 1 and pillar 2.
      • From 2 to 3, 4, 5, 6, 7, 8, 9, 10, 11B and 12
      • From 3 to 12, 11B, 10, 9, 8, 7, 6, 5 and 4

    6.2 Kalgoorlie Baseline

    • From 1 to 2, 3, 4, 5, 6, 7 and 8
    • From 2 to 8, 7, 6, 5, 4 and 3

    6.3 Busselton Baseline

    • From 1 to 2, 3, 4, 5 and 6
    • From 2 to 6, 5, 4 and 3
    • From 3 to 4, 5 and 6
    • From 4 to 6 and 5
    • From 5 to 6

7. Medjil Procedures

  • Step 1: To start a new Staff Calibration, click on the EDM Calibration > EDMI Calibration.

  • Step 2: Click on the Start new calibration button.
    • Site: select the site from the select the box Auto select corresponding calibration of this baseline and select the specfic calibration of the baseline that is to be used for certified distances.
    • Survey date: select date the measurements were observed.
    • : select date or leave as default.
    • Observer: Enter the name of the observer.
    • Weather: Select an applicable description of the weather from the dropdown.
    • Job Number/Reference: Optional identification reference
    • Comment: Optional Comment
    • EDM: select the EDM or enter a new one by clicking on the + button.
    • EDM: select the Prism or enter a new one by clicking on the + button.
    • If meteorological observations have not been applied to the distance observations that are to be imported, unselect the checkbox Atmospheric corrections applied to EDM data
    • Thermometer: select the Thermometer or enter a new one by clicking on the + button.
    • Barometer: select the Barometer or enter a new one by clicking on the + button.
    • Hygrometer: Optionally, select the Hygrometer or enter a new one by clicking on the + button.
    • If Medjil is not to apply corrections to meterological observations included in the import file, unselect: the checkbox
      • Thermometer calibration corrections applied
      • Barometer calibration corrections applied
      • Hygrometer calibration corrections applied
    • Uncertainty budget: select the Uncertainty budget or enter a new one by clicking on the + button.
    • A-priori scalar: Optional global scale factor for adjusting the estimation of uncertainty of the distance measurements.
    • Rejection Criteria for outlier detection: number of standard deviations to use as a threashold for flagging outliers.
    • If Medjil is not calculate a calibration that tests for cyclic errors, unselect the checkbox Test for cyclic errors
    • Scanned Fieldnotes: Click the Choose File button to select the field record (in pdf format).
    • EDM File (*.csv): Click the Choose File button to select the csv file.
    • Click the Submit button.
    • Note:
      • Form errors will be shown in red text to help correctly fill the form.
      • The EDM File (*.csv) must contain the following following column headings and fields:
        • from_pillar: Pillar name as defined in Medjil calibration sites
        • to_pillar: Pillar name as defined in Medjil calibration sites
        • height_of_instrument: Height in metres
        • height_of_target: Height in metres
        • horizontal_direction(dd): Angle in decimal degrees
        • slope_distance: Distance in metres
        • temperature: Temperature in degrees Celsius
        • pressure: Pressure in Hectopascals or millibars
        • humidity: Humidity in percentage
      • Refer to this sample dataset with the corresponding Fieldnotes for the required data format.

8. Potential error sources

There are a number of sources of error inherent in surveying equipment. It is beyond the scope of this procedure to identify all sources of error inherent in surveying equipment. Surveyors should be aware of the limitations of the equipment they use and ensure that it is well maintained and regularly checked.

Please refer to the technical manual 3.1 Systematic errors in EDM instruments for further explaination.

9. Reduction and Interpretation

Medjil has been developed by Landgate for the calibration of EDM instruments against baselines. The calibration determines the instrument constants, errors and their associated uncertainties from field observations for individual EDM instruments. The adjustment is performed according to ISO standard. The calibration results and measurements for each EDM instrument and baseline are stored in a database for future reference and legal traceability.

An EDM instrument calibration generates an EDM Calibration Report, which includes the following sections:

  • Instruments' Details
  • Baseline Details
  • Certified Baseline Distances Used
  • Least Squares Adjustment Summary
  • Estimated Instrument Correction
  • Statistical Tests
  • Calibration Distance Observations
  • Uncertainty Budget - Uncertainty Sources
  • Combined Uncertainty Budget - Total Distance Uncertainty
  • Uncertainty Groups Contribution to Certified Distances
  • Table of Observations, Corrections, Uncertainties and Residuals
  • Residuals From Least Squares Fit of The Baseline Measurements
  • Comparison of Raw Observations to Certified Slope Distances
  • Return Phase Angle of Observed Distances
  • EDMI Calibration History
  • Report Notes
  • Data Warnings
  • Approvals - signature block

The program also generates a concise EDM Calibration Certificate. The Calibration Certificate is an extract of the report and contains the following:

  • Instruments' Details
  • Baseline Details
  • Certified Baseline Distances Used
  • Least Squares Adjustment Summary
  • Estimated Instrument Correction
  • Comparison of Raw Observations to Certified Slope Distances
  • EDMI Calibration History
  • Report Notes
  • Data Warnings
  • Approvals - signature block

Where the statistical analysis reveals the calibration to be outside tolerances, adherence to the test method equipment settings and observation data shall be reviewed to determine the source of the inaccurate results. If changes are made, the calibration shall be re-run and further reviewed. If the statistical analysis continues to fall outside of the tolerance, then the calibration is deemed to be nonconforming.

The following is a suggestion for Interpretation of EDMI calibration reports:

  • Review that the header information is correct for the calibration performed.
  • Instruments' Details - Review the details and confirm that it is correct for the instrumentation used for the calibration.
  • Baseline Details - Check that the date of baseline calibration is appropriate for the EDMI calibration performed.
  • Least Squares Adjustment Summary
    • The chi-square (χ2) test on the variance factor will fail if the a-posteriori variance factor is too high, indicating that larger than expected least squares residuals are present.
    • The chi-square (χ2) test on the variance factor will also fail if the a-posteriori variance factor is too low. This indicates that the weighting of the observations may have been more stringent, with the failure being caused by the a-posteriori variance factor being ‘too good’.
    • When the least squares adjustment fails the chi-square (χ2) test on the variance factor, the uncertainty budget and uncertainty sources should be reviewed.
  • Estimated Instrument Correction - Verify that the Estimated Calibration Parameters are commensurate with the model of EDM under test.
    • As a guide, the estimated Index parameter should not exceed ± a few millimetres.
    • The estimated Scale parameter should not exceed ±3 – 4ppm. Note that a large scale error is often the result of poor modelling of the meteorological conditions.
    • If the EDM has previously undergone calibration, check if these estimated parameters are of a similar order of magnitude.
  • Calibration Distance Observations - For each inter-pillar distance, check:
    • The observed distances are entered correctly.
    • The temperature and atmospheric pressure are entered correctly.
    • A minimum of four distances have been entered.
    • The standard deviation is small (usually sub-millimetre).
  • Table of Observations, Corrections, Uncertainties and Residuals
    • Check the Difference to Certified (mm). The Average Distance values are the unadjusted distance observations corrected only for the first velocity correction.
      • A larger than expected Difference to Certified (mm) may indicate potential pillar movement, especially if it is repeated a second time to that pillar.
      • A larger than expected Difference to Certified (mm) may also indicate an error has occurred entering any one of the field observations (i.e. distance, temperature, pressure, instrument height or target height).
    • Check the Atmospheric Correction has been applied correctly according to any pre-processing completed on the imported raw observations.
    • Check the Uncertainties Total should be should be commensurate with the instrument under calibration.
    • Check the size of the residuals and standard residuals
      • Gross and systematic errors are indicated by the residuals. Values should be commensurate with the instrument specifications.
      • Observations with very large residuals should be investigated and rejected as appropriate.
      • Smaller standardised residuals maybe caused by large random errors or systematic errors.
  • Residuals From Least Squares Fit of The Baseline Measurements - Inter-pillar distances containing standardised residuals which exceed the specified residual rejection criterion (i.e. outliers) are flagged with red formatting. Hover over the data point to identify the inter-pillar distance represented and further investigate the origin of the high standard residual.
  • Comparison of Raw Observations to Certified Slope Distances - Check data to identify unexpected trends and calibration errors.
  • Data Warnings - Check warnings for unexpected comments.
  • Approvals - Following the above-mentioned evaluation of the EDMI Calibration report, and when the analysis indicates the calibration is within tolerance, the approvals signature block can be signed by the surveyor. The report must be saved to record the calibration in the database.

10. Definitions

National Measurement Institute (NMI) - is a Commonwealth statutory authority established in 2004, operating under the National Measurement Act 1960. The Institute is responsible for advising the Government on the scientific, technical and legislative requirements of Australia's national measurement system and has specific responsibilities for co-ordinating the national measurement system for legal metrology.

Verifying Authority - the NMI appoints verifying authorities under Regulation 71 and 73 of the National Measurement Regulations 1999 in accordance with the National Measurement Act 1960. Verifying authorities are appointed where there is a need for legally traceable measurement.

Legally Traceable Measurement - under the National Measurement Act 1960 measurements made for a legal purpose must be able to be traced back to the National Reference Standard.

Legal Metrology - comprises all measurement carried out for legal purpose. It includes all measurement that is subject to regulation, by law or government decree.

Least Square Adjustment - is rigorous method of adjusting redundant measurements which produces the best adjusted values and associated uncertainties.

Please login or signup to manage and calibrate your staff.

or

Barcode Staff Calibration User Guide

  1. 1. Introduction
  2. 2. About the Boya Staff Calibration Range
  3. 3. How to calibrate barcode staves
    1. 3.1 Field Procedures
    2. 3.2 Sample data
    3. 3.3 Medjil Procedures
  4. 4. Conclusion


1. Introduction

Digital levelling systems have long been used by the surveying and engineering industry to determine height differences between any two or multiple locations for various applications. In Western Australia, Landgate is responsible for the national standard of measure for length, in this case, defined by the height difference or length measured in the vertical plane. In an effort to provide an accurate and a uniform system of levelling across the State, Landgate established a staff calibration range at Boya in 2002. The range comprises 2 observing pillars and 21 pins set in a granite outcrop. The pins have been placed at optimum distances from the pillars and cater for testing over a 4 metre height difference. The relative height differences between the pins have been accurately determined by repeat observations using precision levels in conjunction with calibrated invar staves.

The Boya Barcode Staff Calibration Range is monitored on a regular basis. Medjil is an online platform developed by Landgate to ensure that the digital levelling systems, in particular, the barcoded staves are calibrated from time to time maintain its required accuracy. Please go through the following documents before commencing the field procedure:

2. About the Boya Staff Calibration Range

The Landgate barcode staff calibration range is located at the Land Surveyors Licensing Board's examination site at Boya and consists of 2 observing pillars and a series of 21 stainless steel pins set in a solid granite outcrop in a semi-arc rounding the two observing pillars.

The two observing pillars were first constructed beside a large piece of sloping granite which had the required 4 metres of height difference between top and bottom. The highest pillar is set at a comfortable observing height and the lowest a metre lower and closer to the rock and range.

The pins were glued into drilled holes in the granite while the observing pillars were concreted deep into the ground to ensure their stability. With the 3 metre invar staff, it was possible to observe from Pin 1 to 15 from the high pillar and from Pin 7 to Pin 21 from the lower pillar

Observing Pillars

Digital level set on Pillar MV 83 (high pillar)

The low pillar or Pillar B can be seen just below the high pillar..

Invar staff

Invar staff set on Pin 2

Invar staves set up on the pins are levelled and is held firmly by a bipole to main stability during the course of reading.

Pin configuration

Stainless steel pins glued in granite rock

The pins were glued into drilled holes in the granite outcrop in an arc shape with a distance of about 10 metres from the high pillar and 8.8 metres from the low one.

Obsering from Pillar A

Observing from Pillar MV 83(high pillar).

With a 3 metre staff, readings can be done only for the first 15 pins using the high pillar.

3. How to calibrate barcode staves

3.1 Field Procedures

  1. Use the barcode staff booking sheet to record the collimation details and range observations details in the field.
  2. Check the bubble of the digital level and the staff. Re-adjust them if necessary. The staff needs to be truly vertical for the best results.
  3. Check if the Digital Level has an existing staff calibration. If so, delete it or disconnect it from this calibration session.
  4. Perform a collimation test on the digital level and use its software to compute and store the constants calculated. For this purpose, four co-linear spikes in concrete 20 m apart have been established adjacent to the range (see the Access Sketch)
  5. Configure the settings in the digital level to record mean and standard deviation of measurement from multiple readings (at least five), if possible. Otherwise, if the instrument does not allow means from multiple observations, a minimum of three complete sets of observations (see 8 – 10 below) is required for redundancy
  6. Set the digital level on the high observing pillar (MV 83) and shade it from the sun. Pillar B (low pillar) is utilised by Landgate when calibrating the range.
  7. Position staff on the highest pin (Pin No 1) and allow at least 5 minutes to settle and adjust to the ambient air temperature. Record the air temperature in the booking sheet with a shaded thermometer.
  8. A set of observations for a standard four metre staff consist of:
    • Set up on Pillar MV 83
      • backsight to the staff at Pin Number 1
      • intermediate sights to the staff at Pin Numbers 2 to 20
      • foresight to the staff at Pin Number 21
  9. A set of observations for a standard three metre staff consist of:
    • Setup on Pillar MV83
      • backsight to the staff at Pin Number 1
      • intermediate sights to the staff at Pin Numbers 2 to 14
      • foresight to the staff at Pin Number 15
    • Setup on Pillar B
      • backsight to the staff at Pin Number 7
      • intermediate sights to the staff at Pin Numbers 8 to 20
      • foresight to the staff at Pin Number 21
  10. Record the air temperature in the booking sheet at the end of each survey.
  11. Download the level data and copy it in a csv file in the correct format (see Section 3.2) without any headers.
  12. Click here to process your staff readings and print the calibration report in a pdf format.

3.2 Sample data

Data should be uploaded in a correctly formatted text or csv file as per the file format. File Format - All fields are separated by a comma (,) and must NOT contain the header line.

  • Field 1 = Pin number
  • Field 2 = Staff reading
  • Field 3 = Number of readings
  • Field 4 = Standard deviations

Example of a text/csv file. The file can also be downloaded from here.

Pin number, Staff reading, Number of readings, Standard deviations
1, 0.07417, 10, 0.000090
2, 0.16503, 10, 0.000070
3, 0.32868, 10, 0.000020
4, 0.47489, 10, 0.000090
5, 0.68799, 10, 0.000070
6, 0.87413, 10, 0.000040
9, 1.52628, 10, 0.000030
7, 1.07422, 10, 0.000030
8, 1.27955, 10, 0.000050
10, 1.79376, 10, 0.000040
11, 2.12703, 10, 0.000080
12, 2.3912, 10, 0.000050
13, 2.51647, 10, 0.000110
14, 2.64608, 10, 0.000050
15, 2.87513, 10, 0.000120
16, 3.05758, 10, 0.000060
17, 3.20978, 10, 0.000040
18, 3.36207, 10, 0.000120
19, 3.52057, 10, 0.000200
20, 3.6692, 10, 0.000070
21, 3.87885, 10, 0.000100

Other information required are the staff and observation metadata which is recorded in the booking sheet at the time of observation. Retain a copy of the field booking sheet at least until a new calibration is done for the staff

3.3 Medjil Procedures

  • Step 1: To start a new Staff Calibration, click on the Staff Calibration > Staff Calibration.

  • Step 2: Click on the Start new calibration button.
    • Job Number: enter a Job Number with a ten digit alphanumeric code.
    • Calibration Site: select Boya from the dropdown.
    • Staff Number: select the staff number or enter a new one by clicking on the + button. When entering (or creating) a new staff, a new window will pop up and users are advised to enter all the known fields - Make, Model, Owner, Staff Number, Type, Length and CoE. If the staff has been previously calibrated, users can tick the Is Calibrated box. A new window will appear to provide information about the calibration details.
    • Level Number: select the digital level (number) or enter a new one by clicking on the + button.
    • Start temperature: Temperature at the beginning of measurement.
    • End temperature: Temperature at the end of measurement.
    • Field Data: Click the Choose File button to select the csv file. A sample file is described here for reference.
    • Field Book: Click the Choose File button to selec the field book (in pdf format).
    • Enter an Observer name or tick the I am the Observer, if the observer is same as the person performing this procedure.
    • Calibration date: Choose a calibration date.
    • Click the Submit button.
    • Note:
      • Form errors will be shown in red text to help correctly fill the form.
      • Test data is provided here with the corresponding Field Book to assist with the Staff Calibration procedure.

  • Medjil Steps 1-2

  • Step 2: By submitting the form in Step 1, the files will be read and processed to calibrate the staff. The Staff Calibration Report will be displayed in the next window - tabulating the Correction Factor (a multiplicative scale factor) and the graduation uncertainty at 95% confidence level at 25$^\circ$C, staff readings and corrections, the correction factors/errors at various tempeartures. If the temperature exceeds +55$^\circ$C or -10$^\circ$C at Correction Factor = 1, users are advised to check for possible errors in metadata information provided. Wooden staves generally exceed this limit and are generally used for high precision levelling..

    Click on the Print Report >> to print in a pdf format. The report has three pages. A formula is provided in Page 1 just below the Correction Factor on how to apply it to the future height differences (see below).


  • Staff Calibration report

  • Step 3: Staff Calibration Record.

    The staff calibration records are automatically stored in a database. Calibration reports can be retrieved from the Levelling Staff Registry under the Staff Calibration tab.

    Users are also able to add previous calibration record (s) here by clicking the Add new calibration certificate. However, the form only accepts staffs calibrated at Boya Staff Calibration Range and it is also important to note that previous calibration has no impact on the current calibration as they are indpendent of each other.


  • Staff Calibration Record Form

4. Conclusion

The Boya Staff Calibration Range and software developed by the then Department of Land Information (DLI) has enabled surveyors who use digital levels and barcode staves to calibrate their staves in a simple and cost effective manner for use in geodetic and other high order levelling. Both the calibration of the Range and Staves are based on the methods of least squares estimation, which is considered mathematically more rigorous. Medjil will update the estimated (monthly) average of the Staff Calibration Range whenever a new range measurement is added by Landgate. These (monthly) Range values are then used as a reference to calibrate other staves.

Please contact Landgate at geodesy@landgate.wa.gov.au, if you encounter any issues when calibrating a staff. And users are advised to go through this user guide before contacting Landgate.


Copyright © 2020-2024 Western Australian Land Information Authority

Last updated: 24 May 2024